ACWD Assy

8-2-32-1
Tools
Common pilIips screwdrivers
8-2-32-2
Preparations
Shut down the system and switch off the main breaker.
8-2-32-3
Removal procedure
1.) Remove the Side Right Cover. Refer to the
2.) Remove the EMI Cover R. Refer to the
3.) Remove the ASIG Board. Refer to the
4.) Pull the ACWD assy out from Nest box. The ACWD assy is located in slot #2.
8-80
5
Figure 8-115 PWAs in Nest Box
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-16 "Side Right Cover" on page
8-2-27 "EMI Cover R" on page
8-2-35 "ASIG Assy" on page