GE H
EALTHCARE
D
5394141, R
IRECTION
EVISION
8-2-21
Top Bottom Cover
8-2-21-1
Tools
•
Common pilIips screwdrivers
•
Stubby screwdriver (Flat tip and Cross tip)
8-2-21-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-21-3
Removal procedure
1.) Remove the Dummy Cover L/R. Refer to the
2.) Unscrew 7 screws (1-7) from the bottom of the top cover.
3.) Remove the Top Bottom Cover.
4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-22
Service Manual
Section
Section 4-3-1
Section 4-3-2
Section 4-8-1
Section 10-5-5
8-2-21-4
Mounting Procedure
Install the new parts in the reverse order of removal.
5
Figure 8-96 Top bottom cover
Functional Tests
Functional Test / Diagnostic Test
Power On/Boot Up
Power Off / Shutdown
Cover Parts Function Validation
Physical Inspection
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
LOGIQ™ P5 S
8-2-25 "Dummy Cover L / R" on page
"Service Manual, Direction
5394141, Rev 1+, Section 8-2-21. Equipment
passed all required tests and is ready for use. "
M
EVICE
ANUAL
8-66.
Debrief Script
8-61