15" OSD Key Assy Function Validation Procedure

1.) Turn on the system.
2.) Select "OSD Button" and "Lamp Button" marked below in the figure below.
3.) OSD Menu should be displayed within 5sec and the Lamp should be turned on as soon as Lamp
Button is pushed.
4.) Select "OSD Button" marked below in the figure below and check that contrast value is changed as
soon as the button pushed.
5) Displayed value should be changed as soon as the button pushed.
4-28
5
Lamp Button
Figure 4-24 Select OSD Button & Lamp Button
OSD button for adjusting values
Figure 4-25 Select OSD Button
Section 4-6 - Monitor Function Checks