Aps/Aps Pro Assy - GE LOGIQ P5 Service Manual

Hide thumbs Also See for LOGIQ P5:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5394141, R
IRECTION
EVISION
7-6-4

APS/APS Pro Assy

NOTE:
HVL+ and HVL- are not available for APS PRO.
The APS/APS Pro Assy is located at the No. 4 slot of the NEST Assy.
Table 7-11
LED
DP5 (Top)
LP5
AP6
DP33
SHV+
HVL+
HVH+ (Bottom)
7-22
5
LEDs on APS/APS Pro Assy
Descriptions
Digital 5V output indicator
Live Power 5V output indicator
Analog 6V output indicator
Digital 3.3V output indicator
SHV+ output indicator status indicator
HVL+ level indicator
HVH+ level indicator
Section 7-6 - LED Descriptions
LOGIQ™ P5 S
Normal State
Normally On
Normally On
Normally On
Normally On
Normally On
Normally On in CFM mode or PW mode
Normally On in B mode
M
EVICE
ANUAL

Advertisement

Table of Contents
loading