Standard Confirmation & Functional Checks - GE LOGIQ P5 Service Manual

Hide thumbs Also See for LOGIQ P5:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5394141, R
IRECTION
EVISION
8-4-4
Standard Confirmation & Functional Checks
Configure the system settings and perform functional checks after installation of software.
Items for configuration and functional checks include:
-
-
-
-
8-4-4-1
Confirming software version
1.) Touch the Utility button on the Touch Panel
2.) Click on About tab. The software version will be shown on the monitor as shown. Confirm the
following: Refer to the Figure 8-202.
3.) Make sure that Application Software Version (R4.x.x or R3.0.x) ), Software Part Number (described
on the Application Software CD-ROM) and Image Part Number (described on the Base Image
Software CD-ROM).
5
Confirmation of the software version
Preset Data (if necessary)
Patient Image Data (if necessary)
Functional Checks for Service Platform (Diagnostic Program)
Figure 8-202 Confirmation of software version