GE H
EALTHCARE
D
5394141, R
IRECTION
EVISION
8-2-3
OSD button and lamp set, LCD Plastic Filter
8-2-3-1
Tools
•
Common pilIips screwdrivers
8-2-3-2
Preparations
•
Shut down the system and switch off the main breaker.
•
Maneuver control console to a suitable position for removing the monitor.
8-2-3-3
Removal procedure
1.) Unscrew 4 screws (1-4) from the LCD front cover.
2.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-4
Service Manual
Section
Section 4-3-1
Section 4-3-2
Section 4-6-2
Section 10-5-5
8-2-3-4
Mounting Procedure
Install the new parts in the reverse order of removal.
5
(2)
(1)
Figure 8-12 Unscrews 4 screws
Functional Tests
Functional Test / Diagnostic Test
Power On/Boot Up
Power Off / Shutdown
15" OSD Key Assy Function Validation
Physical Inspection
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
(3)
(4)
5394141, Rev 1+, Section 8-2-3. Equipment
passed all required tests and is ready for use. "
Procedure
LOGIQ™ P5 S
M
EVICE
ANUAL
Debrief Script
"Service Manual, Direction
8-9