Diagnostics

NOTE:
When Service Login for FRU Diagnostic, scan mode must be B-mode.
Diagnostics should be performed with probes which have 128 or more elements.
WARNING
WARNING
Detailed Diagnostic information is found in Diagnostics/Troubleshooting.
NOTE:
After finishing Diagnostic test, the system should be rebooted.
5-10-5-1
Diagnostics Execution
Diagnostic tests are executable by both local and remote users. The Service Platform provides top-level
diagnostic selection based on the user's level and login access permissions. Remote access will require
disruptive diagnostic permissions to run Acquisition diagnostics.
5-10-5-2
Diagnostic Reports
Diagnostic tests return a report to the Service Platform. The platform retains the report and allows for
future viewing of the diagnostic logs.
5-10-5-3
Proactive Diagnostics
A system of self-monitoring is largely supported with the integration of iLinq . The scheduler, executive,
user interfaces, and some of the proactive diagnostic functions are provided by iLinq . Other tasks
should be provided by the product team.
5
Figure 5-43 GE Service Diagnostic Page
Section 5-10 - Service Platform
LOGIQ™ P5 S
M
EVICE
ANUAL
5-45