Board Function Checks; Cl1Trx Assy / Bl1Trx Assy Function Validation Procedure - GE LOGIQ P5 Service Manual

Hide thumbs Also See for LOGIQ P5:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5394141, R
IRECTION
EVISION
Section 4-9Board Function Checks
4-9-1

CL1TRX Assy / BL1TRX Assy Function Validation Procedure

When Service Login for FRU Diagnostic, scan mode must be B-mode. After Diagnostic is completed,
WARNING
WARNING
the system should be rebooted.
Diagnostics should be performed with probes which have 128 or more elements.
WARNING
WARNING
1.) Turn on the system and check the completion of Echoloader loading.
2.) Do the Service Login. Refer to the
3.) Select 'Diagnostics' => 'LOGIQ P5 Diagnostics' => 'FRU Test' => 'CL1TRX Assy / BL1TRX overall'.
4.) Execute the diagnostic test.
5.) 5)All diagnostic test item of CL1TRX Assy / BL1TRX should be passed.
6.) Shut down the system.
7.) Restart the system
8.) Check the basic function of system's different modes. Refer to
on page 4-8
5
5-10-2-2 "Service Login" on page
Figure 4-38 CL1TRX Diagnostic test
and refer to
4-3-5 "System CFM and PWD Checks" on page
Section 4-9 - Board Function Checks
LOGIQ™ P5 S
EVICE
5-41.
4-3-4 "System B/M-Mode Checks"
4-9.
M
ANUAL
4-41

Advertisement

Table of Contents
loading

Table of Contents