Before Starting Software Loading - GE LOGIQ P5 Service Manual

Hide thumbs Also See for LOGIQ P5:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5394141, R
IRECTION
EVISION
8-3-5

Before Starting Software Loading

8-3-5-1
Saving Connectivity
Before starting upgrade, write down the Connectivity Setting for back-up.
1.) Go to Utility > Connectivity and click the Tcpip tab.
2.) Write down the following parameters:.
Parameters
Computer Name
Enable DHCP
IP-Address
Subnet Mask
Default Gateway
Note : For a stand-alone system, preset values of IP address, subnet mask, and default gateway, shown
above, must entered.
5
Descriptions (Entry for Stand-Alone System)
Figure 8-159 Connectivity
Section 8-3 - Software Loading Procedure
LOGIQ™ P5 S
M
EVICE
ANUAL
8-121