8-2-5

Monitor Assy

8-2-5-1
Tools
Common pilIips screwdrivers
Allen/Unbraco wrench
8-2-5-2
Preparations
Shut down the system and switch off the main breaker.
Maneuver control console to a suitable position for removing the monitor.
8-2-5-3
Removal procedure
1.) Detach the monitor assy from arm.
-
-
-
-
-
While unscrew the hinge hexa screw from the arm, keep hold the monitor to prevent drop down.
2.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-6
Service Manual
Section
Section 4-3-1
Section 4-3-2
Section 4-6-1
Section 10-5-2
8-2-5-4
Mounting Procedure
Install the new parts in the reverse order of removal
8-12
5
Open monitor cable cover
Unscrew a screw of the power connector fixture bracket
Unplug the power plug and DVI connector
Flip down the monitor
Unscrew the 2 hexa screw of the monitor hinge
Figure 8-15 Disassemble Monitor from STD arm
Functional Tests
Functional Test / Diagnostic Test
15" LCDMON II Assy & LCD Panel with filter
Function Validation Procedure
Functional Checks (See Also Chapter 4)
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
Power On/Boot Up
Power Off / Shutdown
LOGIQ™ P5 S
EVICE
Debrief Script
"Service Manual, Direction
5394141, Rev 1+, Section 8-2-5. Equipment
passed all required tests and is ready for use. "