Lcd Std Arm With Cover - GE LOGIQ P5 Service Manual

Hide thumbs Also See for LOGIQ P5:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5394141, R
IRECTION
EVISION
8-2-4

LCD Std Arm with Cover

8-2-4-1
Tools
Common pilIips screwdrivers
Allen/Unbraco wrench
8-2-4-2
Preparations
Shut down the system and switch off the main breaker.
Maneuver control console to a suitable position for removing the monitor.
8-2-4-3
Removal procedure
1.) Remove the Power Cable, DVI Cable.
POWER CABLE
2.) Unscrew 1 screw to remove the LCD arm cover.
8-10
5
Figure 8-13 Removing the Cable
Figure 8-14 Unscrews 1 screw
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
LOGIQ™ P5 S
M
EVICE
ANUAL
DVI CABLE

Advertisement