LCD Display Cover Hinges Replacement Procedure

NOTE:
This section describes the replacement procedure for the LCD Display Cover hinges only
(Part No. S2381805-2; one pair of hinges).
8-2-6-1
Tools
Phillips screwdriver; Loctite 242 glue
8-2-6-2
Time Required
30 minutes
8-2-6-3
Preparations
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-2-6-4
LCD Display Cover Hinges Removal Procedure
1.) Remove the Control Panel and Keyboard, as described in the
Removal Procedure on page 8 -
2.) Disconnect the LCD Flex ribbon cable from the BEP, as shown in
Use a flat screwdriver to carefully release the cable from the connector.
3.) Loosen and remove the 2 screws from each of the LCD hinges, as shown in
1
EVISION
Figure 8-24 LCD Flex Cable Connected to BEP
Chapter 8 - Replacement Procedures