3-7-6

Connecting the Unit to a Power Source

The connection of the Vivid i n/ Vivid q N ultrasound unit to a power source should be performed by a
qualified person who has completed basic Vivid i n/ Vivid q N System User Training. Use only the power
cords, cables and plugs provided by or designated by GE Medical Systems to connect the unit to the
power source.
Verify compliance with all electrical and safety requirements and check the power cord to verify
CAUTION
that it is intact and of hospital-grade before connecting the unit to the power source. Products
equipped with a power source (wall outlet) plug should be connected to the fixed power socket
that has a protective grounding conductor. Never use an adapter or converter to connect with a
power source plug (for example, a three-prong to two-prong converter).
3-7-6-1
Connecting the Ultrasound Unit to the Electrical Outlet
DANGER: To avoid the risk of fire, power to the system must be supplied from a properly rated
outlet. It is recommended to use a dedicated power outlet. The power plug should not, under any
circumstances, be altered to a configuration rated less than that specified for the current.
DO NOT use an extension cord or adaptor plug. Refer to
2-3
for more details.
NOTE:
It is recommended to connect the AC-DC converter to the ultrasound unit first, before connecting it to
the AC wall outlet.
1.) Connect the AC-DC converter to the ultrasound unit.
2.) Connect the AC-DC converter to the wall outlet.
3-62
1
EVISION
Section 3-7 - Completing the Hardware Installation
V
V
IVID I N AND
IVID Q
2-2-5 "Electrical Requirements" on page
N S
M
ERVICE
ANUAL