Bottom Assembly Replacement Procedure - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
8-2-7

Bottom Assembly Replacement Procedure

8-2-7-1
Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the bottom assembly replacement
procedures.
8-2-7-2
Time Required
30 minutes
8-2-7-3
Preparations
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-2-7-4
Bottom Assembly Removal Procedure
1) Remove the battery as described in the
2) Remove the bearing handle as described in the
3) Remove the hard disk as described in the
4) Remove the Control Panel and Keyboard as described in the
Procedure on page 8 -
5) Remove the P&M Board as described in the
page 8 -
6) Remove the TR32 and RFI Boards as described in the
on page 8 -
7) Remove the BEP as described in the
8) Remove the HVPS as described in the
9) Remove the fans as described in the
10) Remove the LCD Display Frame as described in the
page 8 -
11.)Separate the Bottom Assembly from the LCD Display rear cover by loosening and removing the two
screws on each of the left and right hinges (attached to the Rear Cover and Latch Assembly). Refer
to
Figure 8-23 on page
12.)Remove the Bottom Assembly from the system.
13.)Remove the Vivid i n/ Vivid q N system label from the base of the defective Bottom Assembly (peel
off), ready to use on the replacement Bottom Assembly.
8-2-7-5
Bottom Assembly Installation Procedure
1) Adhere the Vivid i n/ Vivid q N system label (previously removed in
the replacement unit.
2.) Return the new Bottom Assembly to the system.
3.) Attach the LCD Display rear cover to the Bottom Assembly, carefully aligning the left and right
hinges (attached to the Rear Cover and Latch Assembly) and returning and fastening the two
screws on each side, as shown in
Note:
Do not overtighten the screws.
4.) Return the LCD Frame as described in the
5) Return the fans as described in the
1
EVISION
9.
30.
33.
27.
8-19.
Chapter 8 - Replacement Procedures
V
Battery Removal Procedure on page 8 -
Bearing Handle Removal Procedure on page 8 -
Hard Disk Removal Procedure on page 8 -
P&M (Probe and MUX) Board Removal Procedure on
TR32 and RFI Boards Removal Procedure
BEP Removal Procedure on page 8 -
HVPS Removal Procedure on page 8 -
Fan Removal Procedure on page 8 -
LCD Display Frame Removal Procedure on
Figure 8-23 on page
8-19.
LCD Display Frame Installation Procedure on page 8 - 28
Fan Installation Procedure on page 8 -
V
N S
IVID I N AND
IVID Q
ERVICE
5.
Control Panel and Keyboard Removal
38.
46.
49.
step 13
above) to the base of
51.
M
ANUAL
3.
7.
8-25