Installing Software Only Or Patch - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
-
-
8-5-8

Installing Software only or Patch

IMPORTANT It is recommended to store partition C:\ by using the Rollback tool (described in the
NOTICE
Software Roll-back Procedure on page 8 -
1) Connect the DVD\CD-RW to the system's USB port.
2) Connect the DVD\CD-RW power supply to the mains power.
3) Place the CD with Patch or Software Only CD into the tray and close.
4) Turn ON the system.
During boot-up, system will prompt to install software or continue booting into regular mode.
5) Click on Install Software and follow the on-screen instructions.
6.) When done, the Installation Complete message is displayed (refer to
7) Move the cursor over the displayed message window and press the SET button.
8.) Press any key to re-boot the system.
The system re-boots into Normal Scanning mode.
9.) Press Config and then About.
The newly-installed software version is now listed under the System Version tab (refer to
Figure 8-130 on page
10.)Remove the CD from the DVD\CD-RW drive and close the tray.
11.)Proceed to perform the following functionality tests:
-
-
-
8-134
1
EVISION
Calibration Tests on page 7 - 25
Full System Test on page 7 - 10
8-133).
Software Configuration Checks on page 4 - 37
Calibration Tests on page 7 - 25
Full System Test on page 7 - 10
Section 8-5 - Software Loading
V
IVID I N AND
139) after full software installation or upgrade.
V
N S
M
IVID Q
ERVICE
ANUAL
Figure 8-129 on page
8-133).

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vivid q nFl000110Fl000090Fq000001Vivid-iVivid-q

Table of Contents