Lower Cover (Rear) Replacement Procedure - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
8-4-7

Lower Cover (Rear) Replacement Procedure

8-4-7-1
Tools
Appropriate Phillips screwdriver.
8-4-7-2
Preparations
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-4-7-3
Lower Cover (Rear) Removal Procedure
Remove six retaining screws of the lower cover (rear) and while supporting the cover with one hand,
gently remove the cover - see
8-4-7-4
Lower Cover (Rear) Installation Procedure
1) Return the cover to the correct position on the rear of the SafeLock Cart - see
2) While holding the cover to prevent it from falling, fasten the retaining screws of the lower cover
(rear).
3) Proceed to perform the following functionality tests:
-
-
1
EVISION
Figure
Figure 8-82 Lower Cover (Rear) Removal
SafeLock Cart - Grounding Continuity on page 10 - 23
SafeLock Cart - Chassis Current Leakage Test on page 10 - 25
Chapter 8 - Replacement Procedures
8-82).
V
V
N S
IVID I N AND
IVID Q
Retaining Screw
Figure
M
ERVICE
ANUAL
8-82.
8-71