Battery Replacement Procedure - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
8-2-2

Battery Replacement Procedure

8-2-2-1
Tools
None required.
8-2-2-2
Time Required
1 minute
8-2-2-3
Preparations
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-2-2-4
Battery Removal Procedure
1) Disconnect the AC adapter and all cables from the Vivid i n/Vivid q N portable ultrasound scanner.
2) Close the Vivid i n/ Vivid q N LCD cover, making sure both front latches are securely fastened.
Latch
3) Carefully turn the Vivid i n/ Vivid q N over, placing it face down.
4.) While holding the Vivid i n/ Vivid q N steady, pull the locking latch away from the battery and slide
the battery sideways out of the battery compartment to remove it completely from the system.
WHEN REMOVING A DEFECTIVE BATTERY, ENSURE THAT IT IS DISPOSED OF IN
NOTE:
ACCORDANCE WITH LOCAL REGULATIONS. ALTERNATIVELY, FORWARD IT TO
1
EVISION
Figure 8-4 Closing Vivid i n/ Vivid q N LCD Cover
Figure 8-5 Sliding the Battery out of the Battery Compartment
Chapter 8 - Replacement Procedures
V
V
N S
IVID I N AND
IVID Q
Latch
M
ERVICE
ANUAL
8-5