Functional Checks; Basic Controls - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
EVISION
Section 4-3

Functional Checks

4-3-1

Basic Controls

4-3-1-1
Bearing Handle Test
Table 4-5
Step
1
2
3
4
5
4-3-1-2
Alphanumeric Keyboard Test
Table 4-6
Step
1
2
3
4
5
1
Bearing Handle Test - Recommended
Task
Disconnect all power supplies and peripherals.
Move the handle to the maximum lowest position
(beneath the system).
Open the screen
Close the screen
Move the handle from the maximum lowest position to
the maximum upper position.
Pick up the system using the handle.
Alphanumeric Keyboard Test - Recommended
Task
Power-up the System in Normal Scanning Mode
Press Config and select Report
Select the Comment Text Tab
Press all Alphanumeric keys one-by-one.
To quickly test every key (each letter in the alphabet),
type the following sentence:
The quick brown fox jumps over the lazy dog
Repeat the test, while holding down the Shift key.
Use the Backspace key to delete the text you typed.
Chapter 4 - General Procedures and Functional Checks
V
V
IVID I N AND
IVID Q
Expected Result(s)
Verify that no peripherals nor power supplies are
connected to the system.
Ensure that the system is standing stable.
Verify that the handle is able to reach the maximum
lowest and maximum highest positions smoothly.
Ensure that the handle is firmly connected to the
system.
Expected Result(s)
As each key is pressed, the corresponding letter is
typed on the screen.
N S
M
ERVICE
ANUAL
4-19