Handrest Replacement Procedure - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
8-4-16

Handrest Replacement Procedure

NOTE:
If required, the main shelf handrest may be replaced separately, as described below.
8-4-16-1
Tools
Appropriate Phillips screwdrivers.
8-4-16-2
Preparations
Shut down the Vivid i n/ Vivid q N ultrasound unit - System Setup.
8-4-16-3
Handrest Removal Procedure
1) Remove all peripherals from the Vivid i n/Vivid q N portable ultrasound scanner.
2) Working from the front of the SafeLock Cart and below the main shelf, loosen and remove the three
screws that secure the handrest to the main shelf - see
Figure 8-102 Handrest Secured to Main Shelf (viewed from below)
3) Remove the handrest.
8-4-16-4
Handrest Installation Procedure
1) Install a replacement handrest on the SafeLock Cart's main shelf, making sure the three screw
holes are aligned correctly.
2) Return and fasten the three screws (previously removed) to secure the handrest in position.
1
EVISION
Chapter 8 - Replacement Procedures
V
V
N S
IVID I N AND
IVID Q
Figure 8-102
below.
Handrest
Securing screws
M
ERVICE
ANUAL
Main Shelf
8-89