Cleaning
10-5-4-1
General Cleaning
Frequent and diligent cleaning of the Vivid i n/ Vivid q N ultrasound unit reduces the risk of spreading
infection from person to person, and also helps to maintain a clean working environment.
CAUTION: When performing Cleaning Procedures, to prevent the risk of system damage, always
observe the following precautions:
•
Use only cleaning materials and solutions (such as Isopropyl Alcohol [IPA]) as
recommended in the procedures described in
•
Do not use any solutions or products not listed in the Vivid i n/Vivid q N User Manual.
•
Do not spray any liquid directly onto the Vivid i n/ Vivid q N covers, LCD Display or keyboard!
•
Do not allow any liquid to drip or seep into the system.
•
Prior to cleaning, turn OFF power to the system.