Hvps Replacement Procedure - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
8-3-5

HVPS Replacement Procedure

8-3-5-1
Tools
Use the appropriate flat and Phillips-type screw drivers, as indicated in the HVPS replacement
procedure.
8-3-5-2
Time Required
15 minutes
8-3-5-3
Preparation
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-3-5-4
HVPS Removal Procedure
1) Make sure the Vivid i n/ Vivid q N LCD Display cover is securely closed and place the scanner face
down. Remove the battery as described in the
2) Remove the bearing handle, as described in the
3) Disconnect each of the two Power Supply cables from the DC Power Supply (HVPS-to-BEP and
HVPS-to-RFI), as shown below.
4) Loosen the five screws that secure the DC Power Supply to the Bottom Assembly, as shown in
Figure 8-56
8-46
1
EVISION
Figure 8-55 Disconnecting the DC Power Supply Cables
below.
Figure 8-56 Loosening the Screws on the DC Power Supply
Section 8-3 - Internal Component Replacement Procedures
V
IVID I N AND
Battery Removal Procedure on page 8 -
Bearing Handle Removal Procedure on page 8 -
V
N S
M
IVID Q
ERVICE
ANUAL