Archiving And Loading Presets - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
4-2-9

Archiving and Loading Presets

NOTE:
Always save presets before any software reload. This action is necessary to ensure that the presets
loaded after the software reload are as up–to–date as possible.
All user presets except changes to Summary, Anatomy, and Biometry pages, can be saved on an
MO disk for reloading on the system.
NOTICE
Presets should not be saved on the same MO disk as images, because the Archive Menu lists the
images but not the presets stored on a MO disk.
4-2-9-1
Archiving Presets to an MO Disk
STEP
TASK
1.
Insert an empty, formatted MO disk
labelled "PRESETS" and marked with
Vivid i n/ Vivid q N's serial number and
the date, into the MO Drive.
For more information about
formatting, see
Formatting
Removable Media on page
Press Config (or F2); select Admin
2.
then select Backup.
3.
Enable the User Defined
Configuration check box.
4.
Select the MOD as the Destination.
5.
Press Backup Now.
6.
Press <Alt> + E on the alphanumeric
keyboard.
1
EVISION
Table 4-3 Saving Presets
4-15.
The Backup window is displayed.
A copy of the Presets is saved on the MO disk.
The MO disk is ejected.