Speaker Assembly Replacement Procedure - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
8-3-7

Speaker Assembly Replacement Procedure

NOTE:
This section describes the replacement procedure for the Speaker Assembly which is supplied as part
of the Fans and Speaker Kit - Part No. 2412826.
8-3-7-1
Tools
Use the appropriate flat and Phillips-type screw drivers, a wire cutter and the appropriate Hex wrench
keys as indicated in the speaker replacement procedure.
8-3-7-2
Time Required
15 minutes
8-3-7-3
Preparation
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-3-7-4
Speaker Assembly Removal Procedure
1) Remove the bearing handle, as described in the
2) Remove the Control Panel and Keyboard, as described in the
Removal Procedure on page 8 -
3) Loosen and remove the two screws on each of the speaker housings (four in total).
4) Disconnect the speaker cables from the fan assembly.
5) Gently lift the speaker assembly towards you and remove from the system.
8-3-7-5
Speaker Assembly Installation Procedure
1) Return a replacement speaker assembly to the original position, ensuring that the two connection
legs of each speaker are facing the two connection legs of the opposite speaker, otherwise an
electrical short may occur.
2) Reconnect the speaker cables to the fan assembly.
3) Return and fasten the two screws on each of the speaker housings (four in total), previously
removed.
4.) Proceed to perform the following functionality tests:
-
-
-
-
-
8-52
1
EVISION
SafeLock Cart - Grounding Continuity on page 10 - 23
is currently mounted on the SafeLock cart
SafeLock Cart - Chassis Current Leakage Test on page 10 - 25
Calibration Tests on page 7 - 25
Full System Test on page 7 - 10
Speakers Tests on page 4 - 23
Section 8-3 - Internal Component Replacement Procedures
Bearing Handle Removal Procedure on page 8 -
9.
V
V
N S
IVID I N AND
IVID Q
Control Panel and Keyboard
- only if the Vivid i n/ Vivid q N system
M
ERVICE
ANUAL