Monitoring Tests; Accessing The Monitoring Diagnostic Test Options - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
Section 7-5

Monitoring Tests

7-5-1

Accessing the Monitoring Diagnostic Test Options

Access the Diagnostic Test window, as described in the Accessing the Front End Diagnostic
Options section, on page 7-25.
7-5-1-1
Full System Monitoring Test
NOTE:
For a description of this test, refer to the information in
IMPORTANT In order to access an individual board test, it is necessary to insert the service dongle from
NOTICE
system boot-up and enter the Service Password.
1.) In the Diagnostic Test window, trackball to Full System Monitoring Test in the list of Monitoring
tests available for selection from the Diagnostic Test Tree (see
The Full System Monitoring Test name is displayed under Execute; all system components to be
tested are highlighted light blue in the Schematic View, as shown below:
2.) As indicated in the Special Instructions area, disconnect any probes currently connected to the
system and make sure that nothing is touching the probe connectors.
3.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
4.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
prefer testing to continue on failure and later you may review the report.
5.) Trackball to the Start button and press Set.
7-86
1
EVISION
Figure 7-57 Full System Monitoring Test
Section 7-5 - Monitoring Tests
V
V
IVID I N AND
IVID Q
Full System Monitoring Test on page 7 -
Figure 7-57
N S
M
ERVICE
ANUAL
13.
below) and press Set.