GE H
EALTHCARE
D
FQ091013, R
IRECTION
8-4-20
Rear Wheel Replacement Procedure
8-4-20-1
Tools
Appropriate flat screw driver and open wrench.
8-4-20-2
Preparations
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-4-20-3
Rear Wheel Removal Procedure
1) Disconnect mains power from the AC adapter.
2) Disconnect the AC adapter and all cables and peripherals from the Vivid i n/Vivid q N portable
ultrasound scanner.
3) Remove the Vivid i n/ Vivid q N and all peripherals from the SafeLock Cart.
4) Carefully lay the SafeLock Cart on its side.
5) Loosen and remove the four bolts that secure the rear wheel to the SafeLock Cart chassis - see
Figure
6) Remove the wheel from the SafeLock Cart.
1
EVISION
8-107.
Figure 8-107 Rear Wheel Bolted to SafeLock Cart Chassis
Chapter 8 - Replacement Procedures
V
V
N S
IVID I N AND
IVID Q
Two rear wheel securing
bolts shown
M
ERVICE
ANUAL
8-95