Ac Distribution Assembly Replacement Procedure - GE Vivid i n Service Manual

Hide thumbs Also See for Vivid i n:
Table of Contents

Advertisement

GE H
EALTHCARE
D
FQ091013, R
IRECTION
8-4-8

AC Distribution Assembly Replacement Procedure

8-4-8-1
Tools
Appropriate flat and Phillips screwdrivers and a wire cutter
8-4-8-2
Preparation
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-4-8-3
AC Distribution Assembly Removal Procedure
1) Turn OFF power to the SafeLock Cart and disconnect the AC power cable from the AC input
assembly at the base of the cart (rear).
2) Remove the Upper Cover (Rear), as described in the
page 8 -
3) Remove the Lower Cover (Rear), as described in the
page 8 -
4) Remove the three retaining screws of the AC Assembly Cover and remove the cover -
below.
The AC Distribution Assembly is now exposed, as shown in
5) Cut all cable ties holding the cables in position.
6) Unscrew the metal cable fastener from the P1 cable connector (on the right) and remove the cable
from its socket - see
7) Disconnect the peripheral power cable from the AC Distribution Assembly - see
8-72
1
EVISION
64.
71.
Figure 8-83 AC Distribution Assembly and Mounting Bracket
Figure
Section 8-4 - SafeLock Cart Components Replacement
8-84.
V
V
IVID I N AND
IVID Q
Upper Cover (Rear) Removal Procedure on
Lower Cover (Rear) Removal Procedure on
Figure
8-84.
Retaining Screw
N S
M
ERVICE
ANUAL
Figure 8-83
Figure
8-84.