Accessing the Full System Diagnostic Options

In 2D-Mode, press <ALT +Config> on the alphanumeric keyboard. The Diagnostic Test window is
displayed, as shown in
7-3-1-1
Full System Test (FE and BEP)
NOTE:
For a description of this test, refer to the
NOTE:
The Full System Test (FE and BEP) performs a full sequence of complete system tests (automatically)
and requires approximately 10 minutes for completion.
NOTICE
1.) In the Diagnostic Test window, trackball to Full System Test (FE and BEP) in the list of tests
available for selection from the Diagnostic Test Tree (see
System Test (FE and BEP) name is displayed under Execute, and all system components are
highlighted blue in the Schematic View, as shown below:
2.) As indicated in the Special Instructions area, make sure the system is calibrated and disconnect
any probe currently connected to the system.
3.) Under the Special Instructions area, activate the Loops checkbox
and select the required number of Loops from the combo box.
4.) Adjacent to On Fail, select either Stop or Continue, as required.
Stopping on failure enables you to review a failed test, the moment it fails. Alternatively, you may
7-20
1
EVISION
(Figure
Figure 7-5 Full System (FE and BEP) Test
Section 7-3 - Performing Full System Diagnostics
7-1).
Full System Test (FE and BEP) on page 7 -
V
V
N S
IVID I N AND
IVID Q
below) and press Set. The Full
Figure 7-5
M
ERVICE
ANUAL
10.

Advertisement