Rear Handle - GE LOGIQ P5 Service Manual

Hide thumbs Also See for LOGIQ P5:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5394141, R
IRECTION
EVISION
8-2-38

Rear Handle

8-2-38-1
Tools
Common pilIips screwdrivers
Allen/Unbraco wrench
8-2-38-2
Preparations
Shut down the system and switch off the main breaker.
8-2-38-3
Removal procedure
1.) Unscrew 1 screw. Remove the Handle Dummy brkt.
2.) Unscrew 2 screws (1-2) from the back of the rear handle.
3.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-39
Service Manual
Section
Section 4-3-1
Section 4-3-2
Section 4-8-1
Section 10-5-5
8-2-38-4
Mounting Procedure
Install the new parts in the reverse order of removal.
5
Figure 8-126 Handle Dummy Bracket
(1)
Figure 8-127 Rear Handle
Functional Tests
Functional Test / Diagnostic Test
Power On/Boot Up
Power Off / Shutdown
Cover Parts Function Validation
Physical Inspection
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
LOGIQ™ P5 S
EVICE
(2)
Debrief Script
"Service Manual, Direction
5394141, Rev 1+, Section 8-2-38. Equipment
passed all required tests and is ready for use. "
M
ANUAL
8-93

Advertisement

Table of Contents
loading

Table of Contents