Cl1Trx Assy / Bl1Trx Assy - GE LOGIQ P5 Service Manual

Hide thumbs Also See for LOGIQ P5:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5394141, R
IRECTION
EVISION
8-2-29

CL1TRX Assy / BL1TRX Assy

8-2-29-1
Tools
Common pilIips screwdrivers
8-2-29-2
Preparations
Shut down the system and switch off the main breaker.
8-2-29-3
Removal procedure
1.) Remove the Side Right Cover. Refer to the
2.) Remove the EMI Cover R. Refer to the
3.) Unscrew 2 screws (1-2) and eject the PCB.
4.) The CL1TRX(BL1TRX) board is located on the most left slot in the nest box.
8-74
5
Figure 8-109 Eject ASIG Assy
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-16 "Side Right Cover" on page
8-2-27 "EMI Cover R" on page
LOGIQ™ P5 S
M
EVICE
ANUAL
8-52.
8-69.