Sysconml Assy Function Validation Procedure - GE LOGIQ P5 Service Manual

Hide thumbs Also See for LOGIQ P5:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5394141, R
IRECTION
EVISION
4-9-4

SYSCONML Assy Function Validation Procedure

4-9-4-1
L1SYSCON diagnostic Test
1.) Turn on the system and check the completion of echoloader loading.
2.) Do the Service Login. Refer to
3.) Select 'Diagnostics' => 'LOGIQ P5 Diagnostics' => 'FRU test' => 'L1SYSCON' => 'L1SYSCON
Overall'.
4.) Execute the diagnostic test.
5.) All diagnostic test item of L1SYSCON should be passed.
6.) Shut down the system.
7.) Restart the system
8.) Check the basic function of system's different modes. Refer to
on page 4-8
4-9-4-2
Peripheral Test
Refer to
5
5-10-2-2 "Service Login" on page
Figure 4-46 L2SYSCON Diagnostic Test
and refer to
4-3-5 "System CFM and PWD Checks" on page
Section 4-5 "Peripheral Checks" on page
Section 4-9 - Board Function Checks
LOGIQ™ P5 S
5-41.
4-3-4 "System B/M-Mode Checks"
4-9.
4-26.
M
EVICE
ANUAL
4-47