Sysconml Assy Function Validation Procedure - GE LOGIQ P6 Service Manual

Hide thumbs Also See for LOGIQ P6:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5245279, R
IRECTION
EVISION
4-9-4

SYSCONML Assy Function Validation Procedure

4-9-4-1
L2SYSCON diagnostic Test
1.) Turn on the system and check the completion of echoloader loading.
2.) Do the Service Login. Refer to the
3.) Select 'Diagnostics' => 'LOGIQ P6 Diagnostics' => 'FRU test' => 'L2SYSCON' => 'L2SYSCON
Overall'.
4.) Execute the diagnostic test.
5.) All diagnostic test item of L2SYSCON should be passed.
6.) Shut down the system.
7.) Restart the system
8.) Check the basic function of system's different modes. Refer to the
Checks" on page 4-6
4-9-4-2
Peripheral Test
Refer to the
3
Figure 4-42 L2SYSCON Diagnostic Test
and refer to the
Section 4-5 "Peripheral Checks" on page
Section 4-9 - Board Function Checks
LOGIQ™ P6/P6 P
5-7-3-2 "Service Login" on page
4-3-5 "System CFM and PWD Checks" on page
4-21.
S
M
RO
ERVICE
5-25.
4-3-4 "System B/M-Mode
4-7.
ANUAL
4 - 41

Advertisement