GE LOGIQ P6 Service Manual page 396

Hide thumbs Also See for LOGIQ P6:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5245279, R
IRECTION
EVISION
8-7-4
Drawer Option installation
1.) This section describes the installation procedure of the drawer option.
2.) Push the drawer from the front to the rear direction. Refer to the Figure 8-280
3.) Unscrew 2 existing screws from the middle cover.
4.) Using the bracket, fix the drawer by screwing 3 screws each of left and right side of the drawer.
Refer to the Figure 8-281.
8 - 176
3
Figure 8-280 Drawer Option Installation
2337572
(Screw, FH M3x6 WHT)
5176744
(Screw, BH M4x20 WHT)
Figure 8-281 Installation of Drawer Option
Section 8-7 - Mechanical Option Installation instruction
LOGIQ™ P6/P6 P
Middle Cover
S
M
RO
ERVICE
ANUAL