GE LOGIQ P6 Service Manual page 301

Hide thumbs Also See for LOGIQ P6:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5245279, R
IRECTION
EVISION
8-2-35
Rear Hook
8-2-35-1
Tools
Common philips screwdrivers
8-2-35-2
Preparations
Shut down the system and switch off the main breaker.
8-2-35-3
Removal procedure
1.) Unscrew 2 screws (1-2) and remove Rear hook.
2.) Unscrew 2 screws (1-2) and remove.
3.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-36
Service Manual
Section 4-3-1
Section 4-3-2
Section 4-8-1
Section 10-5-5
8-2-35-4
Mounting Procedure
Install the new parts in the reverse order of removal.
3
Figure 8-122 Screws on rear hook
Figure 8-123 Screws on the rear hook fix brkt
Functional Tests
Section
Functional Test / Diagnostic Test
Cover Parts Function Validation
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
Power On/Boot Up
Power Off / Shutdown
Physical Inspection
LOGIQ™ P6/P6 P
S
RO
ERVICE
Debrief Script
"Service Manual, Direction
5245279, Rev 3+, Section 8-2-35. Equipment
passed all required tests and is ready for use. "