Removal procedure (cont'd)
4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-45
Service Manual
Section 4-3-1
Section 4-3-2
Section 4-9-12
Section 10-5-2-2
Section 10-5-5
Section 10-6-2
Section 10-6-3
Section 10-6-4
Section 10-6-5
Section 10-6-6
Section 10-6-7
Section 10-6-9
8-2-44-4
Mounting Procedure
Install the new parts in the reverse order of removal.
3
Functional Tests
Section
Functional Test / Diagnostic Test
ACPC II Assy set Function Validation
GEMS Leakage Current Limits
Outlet Test - Wiring Arrangement - USA &
Chassis Leakage Current Test
Isolated Patient Lead (Source) Leakage–Lead
Isolated Patient Lead (Source) Leakage–Lead
Probe Leakage Current Test
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
Power On/Boot Up
Power Off / Shutdown
Procedure
Peripheral/Option Checks
Probe Maintenance
Canada
Grounding Continuity
to Ground
to Lead
LOGIQ™ P6/P6 P
S
RO
ERVICE
Debrief Script
"Service Manual, Direction
5245279, Rev 3+, Section 8-2-44. Equipment
passed all required tests and is ready for use. "