Assembly procedure (cont'd)
3.) Insert the flexible arm into the flexible arm neck pipe. Refer to the figure below.
4.) Screw 3 screws (2373562, FH M4X10 YEL) to fix the flexible arm stopper after adjusting the holes.
Refer to the figure below.
5.) Screw 1 screw (2159625, PH M4X8 W/SP) to install OP bracket. Refer to the figure below.
3
Figure 8-38 Inserting the flexible arm
Figure 8-39 Fixing the flexible arm stopper
Figure 8-40
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
LOGIQ™ P6/P6 P
Installing OP Bracket