GE LOGIQ P6 Service Manual page 287

Hide thumbs Also See for LOGIQ P6:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5245279, R
IRECTION
EVISION
8-2-27
P3RLY/PI128II Assy
8-2-27-1
Tools
Common philips screwdrivers
8-2-27-2
Preparations
Shut down the system and switch off the main breaker.
8-2-27-3
Removal procedure
1.) Remove the side left and right covers. Refer to the
2.) Remove the Front Cover. Refer to the
3.) Remove the RFC Assy. Refer to the
4.) Unscrew 8 screws (1-8) and remove the metal cover. Refer to the Figure 8-107.
5.) Unscrew 6 screws (1~6, white arrow marked). Refer to the Figure 8-108.
3
Figure 8-107 P3RLY/PI128II Cover
Figure 8-108 P3RLY/PI128II Assy
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
LOGIQ™ P6/P6 P
8-2-14 "Right or Left Side Cover" on page
8-2-15 "Front Cover" on page
8-2-32 "RFC Assy" on page
S
M
RO
ERVICE
8-49.
8-77.
ANUAL
8-47..
8 - 67