GE H
EALTHCARE
D
5245279, R
IRECTION
EVISION
8-7-1-3
Removal procedure (cont'd)
6.) Unscrew 2 screws (5177684, HSH M5x20 WHT) to separate the LCD MONITOR and 1 screw
(2159625, PH M4x8 W/SP) to remove the Cable bracket. Refer to the figure below.
NOTE:
While unscrewing the hinge hexa screw from the arm, keep hold of the monitor to prevent
dropping down.
Figure 8-236 Separating the LCD monitor & Removing the Cable bracket
7.) Unscrew 2 screws (2337572, FH M3x6 WHT) to remove the Arm cover L & R. Refer to the figure
below.
3
Figure 8-237 Removing the Arm cover L & Arm cover R
Section 8-7 - Mechanical Option Installation instruction
LOGIQ™ P6/P6 P
S
RO
ERVICE
M
ANUAL
8 - 157