4.) Press "CW" button to start CWD Mode.
5.) Scan in CWD mode (with Virtual Convex enabled).
6.) Press "B" button to scan in "B" mode.
7.) Press Virtual Convex button to disable Virtual Convex if not already disabled.
8.) Press "CW" button to start CWD Mode.
9.) Scan in CWD mode (with Virtual Convex disabled)
10.)Compare the CWD (with Virtual Convex enabled) sensitivity and CWD (with Virtual Convex
disabled).
11.)Pass if two modes have the same sensitivity.
12.)Do the service login. Refer to the 5-7-3-2 " Service login" on page 5-28
13.)Select "Diagnostics" => "LOGIQ P6 Diagnostics" => "FRU Test" => "CWD64" => "CWD64 Overall"
For LP6 PRO, select 'LOGIQ P6 PRO Diagnostics' => 'FRU test' => 'ACWD' => "ACWD
NOTE:
Overall"
14.)After All diagnostic function check of CWD64 should be Passed.Refer to the Table 7-5, "FRU Test,
3
Figure 4-39 CW mode
Figure 4-40 Virtual Convex Disabled.
Section 4-9 - Board Function Checks