Removal procedure (cont'd)
20.)Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-5
Service Manual
Section 4-3-1
Section 4-3-2
Section 4-8-6
Section 10-5-5
8-2-4-4
Assembly procedure
1.) Insert two of the flexible arm stoppers (A+B) to the Power Cable and DVI cable in the Order
illustrated on the figure below. Make sure that chamfered shape of the flexible Arm stopper should
face inside. Refer to the figure below.
Flexible Arm Stopper
2.) Insert oilless bush into the flexible arm neck pipe and insert two cables into the flexible arm neck
pipe. Refer to the figure below.
8 - 20
3
Functional Tests
Section
Functional Test / Diagnostic Test
New articulation arm with cover Function
Figure 8-36 Inserting cables to the flexible arm stopper
Figure 8-37 Inserting the oilless bush
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
Power On/Boot Up
Power Off / Shutdown
Validation Procedure
Physical Inspection
LOGIQ™ P6/P6 P
S
RO
ERVICE
Debrief Script
"Service Manual, Direction
5245279, Rev 3+, Section 8-2-4. Equipment
passed all required tests and is ready for use. "