GE H
EALTHCARE
D
5245279, R
IRECTION
EVISION
8-2-17
Middle Cover
8-2-17-1
Tools
•
Common philips screwdrivers
•
Stubby screwdriver (Flat tip and Cross tip)
8-2-17-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-17-3
Removal procedure
1.) Remove the Left and Right side covers. Refer to the
2.) And remove the Rear Cover. Refer to the
3.) Remove 2 rubber caps and the two screws from the otp of the middle cover of the system. Refer to
the figure below.
4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-18
Service Manual
Section 4-3-1
Section 4-3-2
Section 4-8-1
Section 10-5-5
8-2-17-4
Mounting Procedure
Install the new parts in the reverse order of removal.
8 - 52
3
1
Figure 8-92 2 rubber cap of middle cover
Functional Tests
Section
Functional Test / Diagnostic Test
Cover Parts Function Validation
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-14 "Right or Left Side Cover" on page 8-47
8-2-16 "Rear Cover" on page
2
Power On/Boot Up
Power Off / Shutdown
Physical Inspection
LOGIQ™ P6/P6 P
S
RO
ERVICE
8-51.
Debrief Script
"Service Manual, Direction
5245279, Rev 3+, Section 8-2-17. Equipment
passed all required tests and is ready for use. "