GE LOGIQ P6 Service Manual page 261

Hide thumbs Also See for LOGIQ P6:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5245279, R
IRECTION
EVISION
8-2-9-3
Removal procedure (cont'd)
3.) Unscrew 6 screws from the back side of the A/N key assy. Refer to the Figure 8-78.
4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-10
Service Manual
Section 4-3-1
Section 4-3-2
Section 4-7-4
8-2-9-4
Mounting Procedure
Install the new parts in the reverse order of removal.
3
1
4
Figure 8-78 Screws on the A/N Key
Functional Tests
Section
Functional Test / Diagnostic Test
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
2
5
Power On/Boot Up
Power Off / Shutdown
A/N Key assy validation
LOGIQ™ P6/P6 P
S
RO
ERVICE
3
6
Debrief Script
"Service Manual, Direction
5245279, Rev 3+, Section 8-2-9. Equipment
passed all required tests and is ready for use. "
M
ANUAL
8 - 41