GE LOGIQ P6 Service Manual page 199

Hide thumbs Also See for LOGIQ P6:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5245279, R
IRECTION
EVISION
7-7-4-2
DTRX64II
These programs are provided for testing the DTRX64II assy.
Table 7-3
Menu
Full Test
Assy Revision
Test
USC I/F Test
FEBUS Test
PGC Test
TEST Wave Test
Beamforming
Data Memory
Test
RX Channel Test
TX Channel Test
3
FRU Test, DTRX64II Assy Diagnostics Menu
Performs all of the DTRX64II board tests listed below
Test EEPROM.
Send test pattern to write to EEPROM and read
Displays the board revision of the DTRX64II board
Checks if the DSP in SYSCONML can access the DTRX64IIboards via FEBUS.
Checks if the DSP in SYSCONML can access the DTRX64II Assy via FEBUS.
Sends the test data High, Mid, Low level to PGC Register, then DTRX64II read result
data and verify it.
Sends the test data High, Mid, Low level to Test wave Register, then DTRX64II read
result data and verify it.
Send the command to read / write to beamforming data memory in DTRX64II assy.
Send test wave pattern to the input stage of preamp on DTRX64II, and perform 64
channel scanning, Read raw data, and analyze it.
Send test wave pattern to the DTRX64II assy, and capture a returned echo signal,
Read the data and analyze it.
Note: APSII HV test is requried prior to the TRX TX channel test.
Section 7-7 - Common Diagnostics
LOGIQ™ P6/P6 P
Descriptions
S
M
RO
ERVICE
ANUAL
User
Level
Tools
All
None
All
None
All
None
All
None
All
None
All
None
All
None
GE FE
Only
GE FE
Only
7 - 21

Advertisement