GE LOGIQ P6 Service Manual page 229

Hide thumbs Also See for LOGIQ P6:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5245279, R
IRECTION
EVISION
8-2-2-3
17" LCDMON OSD Key Assy Replace Procedure (cont'd)
6.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-3
Service Manual
Section 4-3-1
Section 4-3-2
Section 4-6-2
Section 10-5-5
8-2-2-4
Mounting Procedure
Install the new parts in the reverse order of removal
3
Functional Tests
Section
Functional Test / Diagnostic Test
17" OSD Key Assy Function Validation
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
Power On/Boot Up
Power Off / Shutdown
Procedure
Physical Inspection
LOGIQ™ P6/P6 P
S
RO
ERVICE
Debrief Script
"Service Manual, Direction
5245279, Rev 3+, Section 8-2-2. Equipment
passed all required tests and is ready for use. "
M
ANUAL
8 - 9

Advertisement

Table of Contents
loading

Table of Contents