GE LOGIQ P6 Service Manual page 259

Hide thumbs Also See for LOGIQ P6:
Table of Contents

Advertisement

GE H
EALTHCARE
D
5245279, R
IRECTION
EVISION
8-2-8
Trackball Assy
8-2-8-1
Tools
Common philips screwdrivers
8-2-8-2
Preparations
Shut down the system and switch off the main breaker.
8-2-8-3
Removal procedure
1.) Remove the Main Keyboard Assy. Refer to the
2.) Unscrew 4 screws (1-4) and disconnect the cable. Refer to the Figure 8-75.
3.) Perform the following functional tests. If all are successful, include the debrief script provided below.
Table 8-9
Service Manual
Section 4-3-1
Section 4-3-2
Section 4-7-3
8-2-8-4
Mounting Procedure
Install the new parts in the reverse order of removal.
3
Figure 8-75 Screws and connector for Trackball
Functional Tests
Section
Functional Test / Diagnostic Test
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-6 "Main Keyboard Assy" on page
Power On/Boot Up
Power Off / Shutdown
Trackball Assy validation
LOGIQ™ P6/P6 P
S
RO
ERVICE
connector
4 screws
Debrief Script
"Service Manual, Direction
5245279, Rev 3+, Section 8-2-8. Equipment
passed all required tests and is ready for use. "
M
ANUAL
8-36.
8 - 39