Cuvette wash and cuvette conditioner usage

CTT and RTT
10% Cuvette Wash
System solutions
Cuvette Conditioner 1:40 with water
NOTE
The total volume of cuvette wash and cuvette conditioner used by the system can vary
slightly from the volumes provided in the tables above. This is normal behavior.
Cleaning and replenishing the dilution bottle
Materials required:
• Deionized Water
• Physiological saline (0.9% NaCl)
Time: 10 minutes
Analyzer mode: READY
NOTE
The dilution bottle may be cleaned when it is refilled, but must be cleaned at least once a
month.
134
Solution
Approximate
Volume Used
During Wash 2
DPP = 3.6 mL
RPP1 = 29 mL
RPP2 = 29 mL
Solution
Dilution by
system
Cuvette Wash
1:10 with water
Number of
Aspirations
DPP = 0.36 mL
RPP1 = 2.9 mL
RPP2 = 2.9 mL
Volume
Volume of
dispensed
undiluted
per WUD
solution
cycle
used per
WUD cycle
600 μL
60 μL
600 μL
15 μL
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
ADVIA 2400 Operator's Guide
Total number of
WUD cycles for
Wash2
1220 (approx)
Total volume
used for
Wash2
undiluted
44 ml
11 ml