ABB MR Series Startup Procedure

ABB MR Series Startup Procedure

Ft-spectroradiometers

Advertisement

MR Series FT-Spectroradiometers
Start-Up Procedure
This manual contains: Information on the unpacking and set-up of the MR Series FT-
Spectroradiometers.
IMZ8947
Revision 3,1 March 2003

Advertisement

Table of Contents
loading

Summary of Contents for ABB MR Series

  • Page 1 MR Series FT-Spectroradiometers Start-Up Procedure This manual contains: Information on the unpacking and set-up of the MR Series FT- Spectroradiometers. IMZ8947 Revision 3,1 March 2003...
  • Page 3 Disclaimer This document contains product specifications and performance statements that may be in conflict with other ABB published literature, such as product flyers and product catalogs. All specifications, product characteristics, and performance statements included in this document are given as suggestive specifications only.
  • Page 5: Table Of Contents

    A B L E O F O N T E N T S INTRODUCTION ..........................1 UNPACKING AND SETUP ......................3 VISUAL INSPECTION ............................. 3 CARRYING CASES LAYOUT......................... 4 HARDWARE STARTUP ....................... 5 CONNECTOR PLATES ............................ 5 INSTALLATION OF THE DSP BOARD IN THE HOST COMPUTER ............10 PUTTING THE BASIC SYSTEM TOGETHER.....................
  • Page 7: Introduction

    1. INTRODUCTION This section is intended for users who have not received the operation training but want to get the instrument up and running. It also contains useful reference information on the set-up of the instrument (hardware and software) and on packing the instrument in order to ship it. This section also contains information on the handling of hygroscopic materials and should be studied by customers having ordered optional hygroscopic optics for their MR system.
  • Page 9: Unpacking And Setup

    This is important for making a claim should the equipment have been damaged during shipment. Please notify the After Sales Service Department at ABB. Before unpacking the system a minimum delay of 12 hours is recommended to allow the equipment to reach thermal equilibrium with the environment.
  • Page 10: Carrying Cases Layout

    COMPUTER MANUALS AND BACKUP SOFTWARE OR SPARE Accessories DATA CABLE 90' INPUT DETECTOR NARROW ANGLE TELESCOPE COLD SOURCE CCD CAMERA POWER SUPPLY CCD CAMERA FUNEL SPARE MANUALS CCD CAMERA MONITOR DETECTOR MEDIUM ANGLE WIDE ANGLE TELESCOPE TELESCOPE MR Series FT-Spectroradiometers...
  • Page 11: Hardware Startup

    3. HARDWARE STARTUP a. CONNECTOR PLATES There is only one connection between the Spectroradiometer and the computer: a 25 pins serial cable for the MR100 or an optical fibre cable for the MR200. One end of the data cable connects to the respective connector plate for each instrument as indicated on the next two figures.
  • Page 12 It is used to show the progress of the power up self test sequence and to indicate faults in the instrument. When the instrument is working correctly the display will show 7, 9 or A depending on MR Series FT-Spectroradiometers...
  • Page 13 the revision of the instrument control code. The following table list the meaning of the different status display states. Instrument status Instruments prior to October 1996 Instruments prior to April 1997 Instruments after April 1997 display code Revision A Revision B Revision C start code start code...
  • Page 14 MR100: The DSP board occupies two ISA slots in the host computer. Only one external connector is available and is used for the connection of the DB-25 serial data cable. The board is displayed on the next two figures. MR Series FT-Spectroradiometers...
  • Page 15 MR200: The DSP board occupies three ISA slots in the host computer. Two connectors are available. On is used for the connection of the fibre optic data cable. The other (DB-25) is an unused RS232 port. This port could be used to receive RS232 signals coming from an IRIG-B time code converter. The board is displayed on the next two figures.
  • Page 16: Installation Of The Dsp Board In The Host Computer

    DSP board, making sure that it is well inserted in the slots. When the computer is not purchased from ABB, the DSP board comes delivered as a separate item. To install it in the computer, please proceed as follows:...
  • Page 17: Putting The Basic System Together

    You must locate two adjacent free slots for the installation of the MR100 DSP, three for the MR200. You may have to re-arrange the location of existing boards to achieve that goal. Some blank holding plates (please refer to A on the above picture) may be present and must be removed. Please keep the screws.
  • Page 18 Connect the data cable at the proper location on the connector plate of the instrument. 10. Connect the other end of the cable into the DSP board located inside the computer. 11. Turn on both the computer and the spectroradiometer. MR Series FT-Spectroradiometers...
  • Page 19 12. Watch the status indicated on the connector plate of the spectroradiometer and check that the correct LED sequence is followed. 13. Start the acquisition software from the computer (please refer to the next section or the software manual) and go into the alignment/status screen where the interferogram and spectral data are displayed in real time. 14.
  • Page 20: Software Startup

    RAM and at least 10 megabytes of disk space. It will also run under Windows 3.x and Windows 9x as a DOS program. If ABB supplied your computer, Research Acquire was preinstalled at the factory. Otherwise, install the software by inserting the distribution diskette into the floppy disk drive and...
  • Page 21 Figure 2 Communication initialisation screen Figure 3 Start-up screen for Research Acquire Start-Up Procedure...
  • Page 22: Editing The Configuration File (Acquire.cfg)

    This setting can be u seful if you have to process data after acquisition without the instrument plugged. This setting will prevent from communication errors on software startup. Note that this setting will disable the Acquire and Instrument menus. MR Series FT-Spectroradiometers...
  • Page 23: External Trigger Address Setting

    iii. External Trigger address setting If a Parallel Input/Output (PIO) board is installed in the computer, set the parameter trig pio address to the proper I/O address in decimal notation. To know which address to use, please refer to the PIO card documentation.
  • Page 25: Handling And Care Of Hygroscopic Materials

    6. HANDLING AND CARE OF HYGROSCOPIC MATERIALS a. INTRODUCTION The window and beamsplitter materials that are frequently used in the field of FT-IR (Fourier Transform InfraRed) spectrometry can be damaged by exposure to moisture in the surrounding environment of the instrument or storage location.
  • Page 26 TlCl KRS-5 KRS-6 Optical material Figure 5-1. Solubility of various optical materials. Solubility of optical materials in H2O Temperature of 25 degrees Celsius CsBr NaBr NaCl Optical material Figure 5-2. Solubility of various alkali halide optical materials. MR Series FT-Spectroradiometers...
  • Page 27: Handling And Care

    c. HANDLING AND CARE The handling and care of the different halide components is not very difficult, but it does require a conscious effort to be made by the user to follow the suggested guidelines. Some suppliers of these optical materials had been contacted and their procedures for handling of such elements have been reviewed and are restated here for clarity.
  • Page 28: Instrumentation

    If the relative humidity is expected to constantly, or frequently, be higher than 35%, then a dry oil free gas purge to the instrument housing should be applied. (Consult your ABB representative on this aspect.) Exposed hygroscopic materials on the instrumentation, such as a window on top plate or at side port location, can be protected by placing desiccant material near the exposed surface and to cover and seal the optical element and desiccant with a non-water-absorbing material.
  • Page 29 Some notes are given here involving relative humidity as this term can be confusing as it represents only a relative measurement of water and not an absolute one, and, so, will vary with conditions encountered. The relative humidity with respect to water is defined as being the ratio of the amount of water that the air actually holds to the amount of water that the air can possibly hold [or in technical terms;...
  • Page 30: Conclusion

    PO Box 610 Colton, California 92324, USA (714) 825-1793 United Desiccants-Gates 6845 Westfield Ave. Pennsauken, NJ 08110-1582, USA (609) 662-6500 Possible source for heat-sealing equipment: National Instrument Co. 4119 Fordleigh Rd. Baltimore, MD 21215, USA (301) 764-0900 MR Series FT-Spectroradiometers...

This manual is also suitable for:

Mr100Mr200Mr100 dsp

Table of Contents