Removing the turret

If you need to remove or replace your turret, use the following instructions for
proper replacement.
1.
Disconnect the injector cable from the GC.
2.
Open the injector door. Unscrew and remove the knurled nut from the top
of the turret. See Figure 9.
3.
Rotate the turret so that the open section faces the back of the tower and
the arrow on top of the turret points directly back (see Figure 9). Push the
stripper arm toward the back, then lift the turret up so that it clears the
center shaft.
4.
To install the replacement turret, align the tab on the motor hub with the
slot in the underside of the turret.
5.
Push the stripper arm to the back of the tower, insert the turret with the
open section facing the back of the tower and the arrow pointing back, and
install the turret with the tab seated in the slot. Seat the turret on the hub,
rotating it gently if necessary until it drops completely into position. The top
of the threaded shaft should be almost level with the top of the turret.
6.
Replace the knurled nut and tighten it firmly (finger-tight).
7.
Close the injector door.
8.
Plug the injector cable into the GC and check for the Ready light.
Note
For increased accuracy of the needle sampling depth, perform the turret
alignment procedure each time the turret is changed. See Turret alignment on
page 16 .
Installation Guide
Removing the turret
15