HP 7683 Installation Manual page 22

Automatic liquid sampler
Table of Contents

Advertisement

Plunger carrier loop
Bearing clip
Syringe carriage
Figure 12 Removing the needle support assembly
4.
With your finger under the shaft near the bearing on the needle support
assembly, pull gently to release the bearing from the bearing clip in the
syringe carriage.
5.
Carefully use the bearing to pull the rod down until you can lift the assembly
out of the syringe carriage.
Caution
Be careful not to pull the assembly by its metal shaft. The shaft is easily bent.
Replacing the needle support assembly
Guide
Syringe
carriage
tracks
Bearing
Installation Guide
Slide
19