3-9-4-1 DICOM Server IP address on the ultrasound
system
(continued)
Vivid T9/Vivid T8 – Basic Service Manual
5795591-100 English Rev. 3
Figure 3-33. Select DicomWorklist > Properties
7. In the Properties dialog, select the IP-address dialog
pull-down menu to select the Worklist Server.
It is not possible to change the setting in the IP-address filed
by editing it.
To change the IP-Address settings, refer to
Remote Archive's Network Information' on page
Enter the DICOM server's AE Title. This entry is case
sensitive and must match exactly.
Enter the DICOM server's Port No.(Port number).
For some DICOM servers, the default Time-out setting
(30) is too low.
Select OK to close the Worklist properties dialog and
save changes.
Figure 3-34. The DicomWorklist Properties Window
Connectivity setup
3-9-2 'Set the
3-44.