Preparing For Setup; Verify Customer Order; Physical Inspection; Emi Protection - GE Vivid T8 Basic Service Manual

Hide thumbs Also See for Vivid T8:
Table of Contents

Advertisement

System Setup

3-5-1 Verify customer order

3-5-2 Physical inspection

3-5-3 EMI protection

3-18

3-5 Preparing for setup

Compare items received by the customer to that which is listed
on the delivery order. Report any items that are missing, back
ordered, or damaged.
Verify that the system arrived intact (visual inspection).
If the system has been damaged, please refer to
transportation' on page i-14
The Vivid T9/Vivid T8 has been designed to minimize the effects
of Electro-Magnetic Interference (EMI). Many of the covers,
shields, and screws are provided primarily to protect the system
from image artifacts caused by this interference. For this reason,
it is imperative that all covers and hardware are installed and
secured before the unit is put into operation.
See
2-2-4 'EMI limitations' on page 2-9
about EMI protection.
in the beginning of this manual.
for more information
Vivid T9/Vivid T8 – Basic Service Manual